ICAR AICE-JRF/SRF Advanced Analytical Techniques & Foodomics: practice questions
ICAR AICE-JRF/SRF (PhD) — Food Processing Technology, Food Safety & Quality, Processing and Food Engineering · India · ICAR through the National Testing Agency
Original practice questions written to the published syllabus · not past papers
10worked here, free
104written to this exam in this subject
12syllabus units
About ICAR AICE-JRF/SRF
ICAR AICE-JRF/SRF (PhD) — Food Processing Technology, Food Safety & Quality, Processing and Food Engineering is a PhD, fellowship and lectureship exam run by ICAR through the National Testing Agency, for candidates in India. Computer-based, multiple choice.
Checked against the exam body's published information. Formats and fees change; confirm the current edition there before you book.
Advanced Analytical Techniques & Foodomics in ICAR AICE-JRF/SRF
The analytical science of a research laboratory: high-resolution mass spectrometry, NMR, advanced chromatography, imaging and hyperspectral methods, proteomics, metabolomics and lipidomics of foods, biosensors, and the chemometrics that interprets the data.
Our 104 ICAR AICE-JRF/SRF questions in this subject: 22 easy, 70 medium, 12 hard, across 12 units.
10 practice questions, worked
Choose an answer, then open the working. Each explanation says why the right answer is right and why the others are not.
These 10 were written to ICAR AICE-JRF/SRF and are not shown on any other exam's page.
Question 1 · easy
Which ionisation technique, used with gas chromatography at 70 eV, gives reproducible fragment spectra searchable in libraries?
Electrospray ionisation
Electron ionisation
MALDI
Atmospheric-pressure photoionisation
Answer and working
B. Electron ionisation
EI at 70 eV gives standardised fragmentation for library matching. ESI is soft and used with LC; MALDI suits large biomolecules.
Question 2 · easy
On a 400 MHz spectrometer, a signal appears 1600 Hz from TMS. What is its chemical shift?
4.0 ppm
40 ppm
0.25 ppm
1.6 ppm
Answer and working
A. 4.0 ppm
δ = 1600 Hz/400 MHz = 4.0 ppm. 0.25 inverts the ratio; 1.6 and 40 slip factors.
Question 3 · medium
What is the main limitation of NMR compared with mass spectrometry for food analysis?
Poor reproducibility between instruments
Inability to quantify
Destruction of the sample
Lower sensitivity
Answer and working
D. Lower sensitivity
NMR is reproducible, quantitative and non-destructive but far less sensitive, so it misses trace compounds.
Question 4 · medium
Why do sub-2 µm particles in UHPLC allow faster separations without losing efficiency?
They increase longitudinal diffusion of analytes
They lower the column back-pressure markedly
They lower the A and C terms of the van Deemter curve
They remove the need for any mobile phase
Answer and working
C. They lower the A and C terms of the van Deemter curve
Smaller particles give lower plate heights and a flatter van Deemter curve at high velocity, at the cost of higher pressure.
Question 5 · easy
What happens to the Maltese cross of starch granules under polarised light on gelatinisation?
It becomes brighter
It disappears
It turns blue
It doubles in number
Answer and working
B. It disappears
Loss of ordered (birefringent) structure on gelatinisation removes the cross.
Question 6 · medium
Which XRD pattern indicates amylose-lipid complexes formed during extrusion?
V-type
A-type
C-type
B-type
Answer and working
A. V-type
Single amylose helices around lipids give the V pattern.
Question 7 · medium
Which wavelength range is used by short-wave infrared hyperspectral cameras?
About 200 to 400 nm
About 5 to 10 µm
About 1000 to 2500 nm
About 400 to 700 nm only
Answer and working
C. About 1000 to 2500 nm
SWIR covers ~1000–2500 nm, where O–H, C–H and N–H overtones occur.
Question 8 · medium
Which pre-processing method reduces light-scattering effects in reflectance spectra?
Mean-centring only
Standard normal variate
Random shuffling of variables
Adding white noise
Answer and working
B. Standard normal variate
SNV (and MSC) correct for scatter differences caused by particle size and surface.
Question 9 · medium
A PLS-DA model separates organic from conventional tomatoes perfectly in training. What must be done before trusting it?
Add more variables until it is perfect
Validate with permutation tests and independent samples
Publish immediately as a definitive method
Remove the cross-validation step
Answer and working
B. Validate with permutation tests and independent samples
With many features and few samples, perfect training separation can be chance; validation shows real predictive power.
Question 10 · easy
What does autoscaling do to each variable?
Sorts values in descending order
Mean-centres it and divides by its standard deviation
Squares every value
Removes all variables with negative values
Answer and working
B. Mean-centres it and divides by its standard deviation
Autoscaling gives each variable unit variance so that all have equal weight.
A study plan for Advanced Analytical Techniques & Foodomics in ICAR AICE-JRF/SRF
This plan follows where our questions actually fall, so the time goes where the marks are likely to be. It is our reading of the syllabus, not the exam body's weighting.
First, the heavy units (35% of the 104 questions): Advanced chromatography: UHPLC, 2D chromatography and ion chromatography (9, 9%); Biosensors and electronic nose and tongue (9, 9%); Chemometrics and machine learning for analytical data (9, 9%); High-resolution and tandem mass spectrometry (9, 9%).
Read the manual chapters for these first and do every worked question you can find on them.
Then the middle band: LC-MS/MS and GC-MS/MS for residues and contaminants (9, 9%); Metabolomics and lipidomics (9, 9%); Proteomics, peptidomics and allergen detection (9, 9%); Thermal analysis: DSC and TGA (9, 9%).
Aim for one sitting per unit, then mixed questions across them.
Then the lighter units: Hyperspectral imaging and machine vision (8, 8%); Microscopy of foods: SEM, TEM, confocal and AFM (8, 8%); Nuclear magnetic resonance in food analysis (8, 8%); X-ray diffraction and scattering (8, 8%).
Learn the definitions, classifications and standard numbers; these units are more often recalled than worked.
Leave time for the hard ones: Hyperspectral imaging and machine vision (1 hard of 8), Microscopy of foods: SEM, TEM, confocal and AFM (1 hard of 8), Nuclear magnetic resonance in food analysis (1 hard of 8).
Come back to these after a first pass, untimed, and read every explanation, including for the options you rejected.
Bank the easy marks: Thermal analysis: DSC and TGA (3 easy of 9); Hyperspectral imaging and machine vision (2 easy of 8).
Do these early in revision so they are fast on the day.
Finish timed. Sit the subject as a timed set without notes, mark it, and send each wrong answer back to its unit above.